<?xml version="1.0"?>
<?xml-stylesheet type="text/xsl" href="cfr.xsl"?>
<CFRGRANULE xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:noNamespaceSchemaLocation="CFRMergedXML.xsd">
  <FDSYS>
    <CFRTITLE>21</CFRTITLE>
    <CFRTITLETEXT>Food and Drugs</CFRTITLETEXT>
    <VOL>8</VOL>
    <DATE>2003-04-01</DATE>
    <ORIGINALDATE>2003-04-01</ORIGINALDATE>
    <COVERONLY>false</COVERONLY>
    <TITLE>Shoulder joint metal/polymer/metal nonconstrained or semi-constrained porous-coated uncemented prosthesis.</TITLE>
    <GRANULENUM>888.3670</GRANULENUM>
    <HEADING>Section 888.3670</HEADING>
    <ANCESTORS>
      <PARENT HEADING="Title 21" SEQ="4">Food and Drugs</PARENT>
      <PARENT HEADING="CHAPTER I" SEQ="3">FOOD AND DRUG ADMINISTRATION, DEPARTMENT OF HEALTH AND HUMAN SERVICES (CONTINUED)</PARENT>
      <PARENT HEADING="SUBCHAPTER H" SEQ="2">MEDICAL DEVICES</PARENT>
      <PARENT HEADING="PART 888" SEQ="1">ORTHOPEDIC DEVICES</PARENT>
      <PARENT HEADING="Subpart D" SEQ="0">Prosthetic Devices</PARENT>
    </ANCESTORS>
  </FDSYS>
  <SECTION>
    <SECTNO>§ 888.3670</SECTNO>
    <SUBJECT>Shoulder joint metal/polymer/metal nonconstrained or semi-constrained porous-coated uncemented prosthesis.</SUBJECT>
    <P>(a) <E T="03">Identification</E>. A shoulder joint metal/polymer/metal nonconstrained or semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits movement in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral component made of alloys such as cobalt-chromium-molybdenum (Co-Cr-Mo) and titanium-aluminum-vanadium (Ti-6Al-4V) alloys, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, or a combination of an articulating ultra-high molecular weight bearing surface fixed in a metal shell made of alloys such as Co-Cr-Mo and Ti-6Al-4V. The humeral component and glenoid backing have a porous coating made of, in the case of Co-Cr-Mo components, beads of the same alloy or commercially pure titanium powder, and in the case of Ti-6Al-4V components, beads or fibers of commercially pure titanium or Ti-6Al-4V alloy, or commercially pure titanium powder. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement.</P>
    <P>(b) <E T="03">Classification</E>. Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance: Shoulder Joint Metal/Polymer/Metal Nonconstrained or Semi-Constrained Porous-Coated Uncemented Prosthesis.”</P>
    <CITA>[66 FR 12737, Feb. 28, 2001]</CITA>
  </SECTION>