[Title 40 CFR 60.485a]
[Code of Federal Regulations (annual edition) - July 1, 2009 Edition]
[Title 40 - PROTECTION OF ENVIRONMENT]
[Chapter I - ENVIRONMENTAL PROTECTION AGENCY (CONTINUED)]
[Subchapter C - AIR PROGRAMS (CONTINUED)]
[Part 60 - STANDARDS OF PERFORMANCE FOR NEW STATIONARY SOURCES--]
[Subpart Vva - Standards of Performance for Equipment Leaks of VOC in the]
[Sec. 60.485a - Test methods and procedures.]
[From the U.S. Government Printing Office]
40PROTECTION OF ENVIRONMENT62009-07-012009-07-01falseTest methods and procedures.60.485aSec. 60.485aPROTECTION OF ENVIRONMENTENVIRONMENTAL PROTECTION AGENCY (CONTINUED)AIR PROGRAMS (CONTINUED)STANDARDS OF PERFORMANCE FOR NEW STATIONARY SOURCES--Standards of Performance for Equipment Leaks of VOC in the
Sec. 60.485a Test methods and procedures.
(a) In conducting the performance tests required in Sec. 60.8, the
owner or operator shall use as reference methods and procedures the test
methods in appendix A of this part or other methods and procedures as
specified in this section, except as provided in Sec. 60.8(b).
(b) The owner or operator shall determine compliance with the
standards in Sec. Sec. 60.482-1a through 60.482-11a, 60.483a, and
60.484a as follows:
(1) Method 21 shall be used to determine the presence of leaking
sources. The instrument shall be calibrated before use each day of its
use by the procedures specified in Method 21 of appendix A-7 of this
part. The following calibration gases shall be used:
(i) Zero air (less than 10 ppm of hydrocarbon in air); and
(ii) A mixture of methane or n-hexane and air at a concentration no
more than 2,000 ppm greater than the leak definition concentration of
the equipment monitored. If the monitoring instrument's design allows
for multiple calibration scales, then the lower scale shall be
calibrated with a calibration gas that is no higher than 2,000 ppm above
the concentration specified as a leak, and the highest scale shall be
calibrated with a calibration gas that is approximately equal to 10,000
ppm. If only one scale on an instrument will be used during monitoring,
the owner or operator need not calibrate the scales that will not be
used during that day's monitoring.
(2) A calibration drift assessment shall be performed, at a minimum,
at the end of each monitoring day. Check the instrument using the same
calibration gas(es) that were used to calibrate the instrument before
use. Follow the procedures specified in Method 21 of appendix A-7 of
this part, Section 10.1, except do not adjust the meter readout to
correspond to the calibration gas value. Record the instrument reading
for each scale used as specified in Sec. 60.486a(e)(7). Calculate the
average algebraic difference between the three meter readings and the
most recent calibration value. Divide this algebraic difference by the
initial calibration value and multiply by 100 to express the calibration
drift as a percentage. If any calibration drift assessment shows a
negative drift of more than 10 percent from the initial calibration
value, then all equipment monitored since the last calibration with
instrument readings below the appropriate leak definition and above the
leak definition multiplied by (100 minus the percent of negative drift/
divided by 100) must be re-monitored. If any calibration drift
assessment shows a positive drift of more than 10 percent from the
initial calibration value, then, at the owner/operator's discretion, all
equipment since the last calibration with instrument readings above the
appropriate leak definition and below the leak definition multiplied by
(100 plus the percent of positive drift/divided by 100) may be re-
monitored.
(c) The owner or operator shall determine compliance with the no-
detectable-emission standards in Sec. Sec. 60.482-2a(e), 60.482-3a(i),
60.482-4a, 60.482-7a(f), and 60.482-10a(e) as follows:
(1) The requirements of paragraph (b) shall apply.
(2) Method 21 of appendix A-7 of this part shall be used to
determine the background level. All potential leak interfaces shall be
traversed as close to the interface as possible. The arithmetic
difference between the maximum concentration indicated by the instrument
and the background level is compared with 500 ppm for determining
compliance.
(d) The owner or operator shall test each piece of equipment unless
he demonstrates that a process unit is not in VOC service, i.e., that
the VOC content would never be reasonably expected to exceed 10 percent
by weight. For purposes of this demonstration, the following methods and
procedures shall be used:
(1) Procedures that conform to the general methods in ASTM E260-73,
91, or 96, E168-67, 77, or 92, E169-63, 77, or 93 (incorporated by
reference--see Sec. 60.17) shall be used to determine the percent
[[Page 540]]
VOC content in the process fluid that is contained in or contacts a
piece of equipment.
(2) Organic compounds that are considered by the Administrator to
have negligible photochemical reactivity may be excluded from the total
quantity of organic compounds in determining the VOC content of the
process fluid.
(3) Engineering judgment may be used to estimate the VOC content, if
a piece of equipment had not been shown previously to be in service. If
the Administrator disagrees with the judgment, paragraphs (d)(1) and (2)
of this section shall be used to resolve the disagreement.
(e) The owner or operator shall demonstrate that a piece of
equipment is in light liquid service by showing that all the following
conditions apply:
(1) The vapor pressure of one or more of the organic components is
greater than 0.3 kPa at 20 [deg]C (1.2 in. H2O at 68 [deg]F).
Standard reference texts or ASTM D2879-83, 96, or 97 (incorporated by
reference--see Sec. 60.17) shall be used to determine the vapor
pressures.
(2) The total concentration of the pure organic components having a
vapor pressure greater than 0.3 kPa at 20 [deg]C (1.2 in. H2O
at 68 [deg]F) is equal to or greater than 20 percent by weight.
(3) The fluid is a liquid at operating conditions.
(f) Samples used in conjunction with paragraphs (d), (e), and (g) of
this section shall be representative of the process fluid that is
contained in or contacts the equipment or the gas being combusted in the
flare.
(g) The owner or operator shall determine compliance with the
standards of flares as follows:
(1) Method 22 of appendix A-7 of this part shall be used to
determine visible emissions.
(2) A thermocouple or any other equivalent device shall be used to
monitor the presence of a pilot flame in the flare.
(3) The maximum permitted velocity for air assisted flares shall be
computed using the following equation:
Vmax = K1 + K2HT
Where:
Vmax = Maximum permitted velocity, m/sec (ft/sec).
HT = Net heating value of the gas being combusted, MJ/scm
(Btu/scf).
K1 = 8.706 m/sec (metric units) = 28.56 ft/sec (English
units).
K2 = 0.7084 m\4\/(MJ-sec) (metric units) = 0.087 ft\4\/(Btu-
sec) (English units).
(4) The net heating value (HT) of the gas being combusted in a flare
shall be computed using the following equation:
[GRAPHIC] [TIFF OMITTED] TR16NO07.038
Where:
K = Conversion constant, 1.740 x 10-\7\ (g-mole)(MJ)/(ppm-
scm-kcal) (metric units) = 4.674 x 10-\6\ [(g-mole)(Btu)/
(ppm-scf-kcal)] (English units).
Ci = Concentration of sample component ``i,'' ppm
Hi = net heat of combustion of sample component ``i'' at 25
[deg]C and 760 mm Hg (77 [deg]F and 14.7 psi), kcal/g-mole.
(5) Method 18 of appendix A-6 of this part or ASTM D6420-99 (2004)
(where the target compound(s) are those listed in Section 1.1 of ASTM
D6420-99, and the target concentration is between 150 parts per billion
by volume and 100 ppmv) and ASTM D2504-67, 77, or 88 (Reapproved 1993)
(incorporated by reference-see Sec. 60.17) shall be used to determine
the concentration of sample component ``i.''
(6) ASTM D2382-76 or 88 or D4809-95 (incorporated by reference-see
Sec. 60.17) shall be used to determine the net heat of combustion of
component ``i'' if published values are not available or cannot be
calculated.
(7) Method 2, 2A, 2C, or 2D of appendix A-7 of this part, as
appropriate, shall be used to determine the actual exit velocity of a
flare. If needed, the unobstructed (free) cross-sectional area of the
flare tip shall be used.
(h) The owner or operator shall determine compliance with Sec.
60.483-1a or Sec. 60.483-2a as follows:
(1) The percent of valves leaking shall be determined using the
following equation:
%VL = (VL / VT) * 100
Where:
%VL = Percent leaking valves.
VL = Number of valves found leaking.
[[Page 541]]
VT = The sum of the total number of valves monitored.
(2) The total number of valves monitored shall include difficult-to-
monitor and unsafe-to-monitor valves only during the monitoring period
in which those valves are monitored.
(3) The number of valves leaking shall include valves for which
repair has been delayed.
(4) Any new valve that is not monitored within 30 days of being
placed in service shall be included in the number of valves leaking and
the total number of valves monitored for the monitoring period in which
the valve is placed in service.
(5) If the process unit has been subdivided in accordance with Sec.
60.482-7a(c)(1)(ii), the sum of valves found leaking during a monitoring
period includes all subgroups.
(6) The total number of valves monitored does not include a valve
monitored to verify repair.