[Title 21 CFR 892]
[Code of Federal Regulations (annual edition) - April 1, 1999 Edition]
[Title 21 - FOOD AND DRUGS]
[Chapter I - FOOD AND DRUG ADMINISTRATION,]
[Subchapter H - MEDICAL DEVICES]
[Part 892 - RADIOLOGY DEVICES]
[From the U.S. Government Printing Office]


21FOOD AND DRUGS81999-04-011999-04-01falseRADIOLOGY DEVICES892PART 892FOOD AND DRUGSFOOD AND DRUG ADMINISTRATION,MEDICAL DEVICES
PART 892--RADIOLOGY DEVICES--Table of Contents




                      Subpart A--General Provisions

Sec.
892.1  Scope.
892.3  Effective dates of requirement for premarket approval.
892.9  Limitations of exemptions from section 510(k) of the Federal 
          Food, Drug, and Cosmetic Act (the act).

                      Subpart B--Diagnostic Devices

892.1000  Magnetic resonance diagnostic device.
892.1100  Scintillation (gamma) camera.
892.1110  Positron camera.
892.1130  Nuclear whole body counter.
892.1170  Bone densitometer.
892.1200  Emission computed tomography system.
892.1220  Fluorescent scanner.
892.1300  Nuclear rectilinear scanner.
892.1310  Nuclear tomography system.
892.1320  Nuclear uptake probe.
892.1330  Nuclear whole body scanner.
892.1350  Nuclear scanning bed.
892.1360  Radionuclide dose calibrator.
892.1370  Nuclear anthropomorphic phantom.
892.1380  Nuclear flood source phantom.
892.1390  Radionuclide rebreathing system.
892.1400  Nuclear sealed calibration source.
892.1410  Nuclear electrocardiograph synchronizer.
892.1420  Radionuclide test pattern phantom.
892.1540  Nonfetal ultrasonic monitor.
892.1550  Ultrasonic pulsed doppler imaging system.
892.1560  Ultrasonic pulsed echo imaging system.
892.1570  Diagnostic ultrasonic transducer.
892.1600  Angiographic x-ray system.
892.1610  Diagnostic x-ray beam-limiting device.
892.1620  Cine or spot fluorographic x-ray camera.
892.1630  Electrostatic x-ray imaging system.
892.1640  Radiographic film marking system.
892.1650  Image-intensified fluoroscopic x-ray system.
892.1660  Non-image-intensified fluoroscopic x-ray system.
892.1670  Spot-film device.
892.1680  Stationary x-ray system.
892.1700  Diagnostic x-ray high voltage generator.
892.1710  Mammographic x-ray system.
892.1720  Mobile x-ray system.
892.1730  Photofluorographic x-ray system.
892.1740  Tomographic x-ray system.
892.1750  Computed tomography x-ray system.
892.1760  Diagnostic x-ray tube housing assembly.
892.1770  Diagnostic x-ray tube mount.
892.1820  Pneumoencephalographic chair.
892.1830  Radiologic patient cradle.
892.1840  Radiographic film.
892.1850  Radiographic film cassette.
892.1860  Radiographic film/cassette changer.
892.1870  Radiographic film/cassette changer programmer.
892.1880  Wall-mounted radiographic cassette holder.
892.1890  Radiographic film illuminator.
892.1900  Automatic radiographic film processor.
892.1910  Radiographic grid.
892.1920  Radiographic head holder.
892.1940  Radiologic quality assurance instrument.
892.1950  Radiographic anthropomorphic phantom.
892.1960  Radiographic intensifying screen.
892.1970  Radiographic ECG/respirator synchronizer.
892.1980  Radiologic table.
892.1990  Transilluminator for breast evaluation.
892.2010  Medical image storage device.
892.2020  Medical image communications device.
892.2030  Medical image digitizer.
892.2040  Medical image hardcopy device.
892.2050  Picture archiving and communications system.

                        Subparts C-E  [Reserved]

                     Subpart F--Therapeutic Devices

892.5050  Medical charged-particle radiation therapy system.
892.5300  Medical neutron radiation therapy system.

[[Page 478]]

892.5650  Manual radionuclide applicator system.
892.5700  Remote controlled radionuclide applicator system.
892.5710  Radiation therapy beam-shaping block.
892.5730  Radionuclide brachytherapy source.
892.5740  Radionuclide teletherapy source.
892.5750  Radionuclide radiation therapy system.
892.5770  Powered radiation therapy patient support assembly.
892.5780  Light beam patient position indicator.
892.5840  Radiation therapy simulation system.
892.5900  X-ray radiation therapy system.
892.5930  Therapeutic x-ray tube housing assembly.

                    Subpart G--Miscellaneous Devices

892.6500  Personnel protective shield.

    Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.

    Source: 53 FR 1567, Jan. 20, 1988, unless otherwise noted.



                      Subpart A--General Provisions



Sec. 892.1  Scope.

    (a) This part sets forth the classification of radiology devices 
intended for human use that are in commercial distribution.
    (b) The identification of a device in a regulation in this part is 
not a precise description of every device that is, or will be, subject 
to the regulation. A manufacturer who submits a premarket notification 
submission for a device under part 807 cannot show merely that the 
device is accurately described by the section title and identification 
provision of a regulation in this part but shall state why the device is 
substantially equivalent to other devices, as required by Sec. 807.87.
    (c) To avoid duplicative listings, a radiology device that has two 
or more types of uses (e.g., use both as a diagnostic device and a 
therapeutic device) is listed in one subpart only.
    (d) References in this part to regulatory sections of the Code of 
Federal Regulations are to chapter I of this title 21, unless otherwise 
noted.



Sec. 892.3  Effective dates of requirement for premarket approval.

    A device included in this part that is classified into class III 
(premarket approval) shall not be commercially distributed after the 
date shown in the regulation classifying the device unless the 
manufacturer has an approval under section 515 of the act (unless an 
exemption has been granted under section 520(g)(2) of the act). An 
approval under section 515 of the act consists of FDA's issuance of an 
order approving an application for premarket approval (PMA) for the 
device or declaring completed a product development protocol (PDP) for 
the device.
    (a) Before FDA requires that a device commercially distributed 
before the enactment date of the amendments, or a device that has been 
found substantially equivalent to such a device, has an approval under 
section 515 of the act, FDA must promulgate a regulation under section 
515(b) of the act requiring such approval, except as provided in 
paragraph (b) of this section. Such a regulation under section 515(b) of 
the act shall not be effective during the grace period ending on the 
90th day after its promulgation or on the last day of the 30th full 
calendar month after the regulation that classifies the device into 
class III is effective, whichever is later. See section 501(f)(2)(B) of 
the act. Accordingly, unless an effective date of the requirement for 
premarket approval is shown in the regulation for a device classified 
into class III in this part, the device may be commercially distributed 
without FDA's issuance of an order approving a PMA or declaring 
completed a PDP for the device. If FDA promulgates a regulation under 
section 515(b) of the act requiring premarket approval for a device, 
section 501(f)(1)(A) of the act applies to the device.
    (b) Any new, not substantially equivalent, device introduced into 
commercial distribution on or after May 28, 1976, including a device 
formerly marketed that has been substantially altered, is classified by 
statute (section 513(f) of the act) into class III without any grace 
period and FDA must have issued an order approving a PMA or declaring 
completed a PDP for the device

[[Page 479]]

before the device is commercially distributed unless it is reclassified. 
If FDA knows that a device being commerically distributed may be a 
``new'' device as defined in this section because of any new intended 
use or other reasons, FDA may codify the statutory classification of the 
device into class III for such new use. Accordingly, the regulation for 
such a class III device states that as of the enactment date of the 
amendments, May 28, 1976, the device must have an approval under section 
515 of the act before commercial distribution.



Sec. 892.9  Limitations of exemptions from section 510(k) of the Federal Food, Drug, and Cosmetic Act (the act).

    (a) The Food and Drug Administration's (FDA's) decision to grant an 
exemption from the requirement of premarket notification (section 510(k) 
of the act) for a generic type of class I device is based upon the 
existing and reasonably foreseeable characteristics of commercially 
distributed devices within that generic type. Because FDA cannot 
anticipate every change in intended use or characteristic that could 
significantly affect a device's safety or effectiveness, manufacturers 
of any commercially distributed class I device for which FDA has granted 
an exemption from the requirement of premarket notification must still 
submit a premarket notification to FDA before introducing or delivering 
for introduction into interstate commerce for commercial distribution 
the device when:
    (1) The device is intended for a use different from its intended use 
before May 28, 1976, or the device is intended for a use different from 
the intended use of a preamendments device to which it had been 
determined to be substantially equivalent; e.g., the device is intended 
for a different medical purpose, or the device is intended for lay use 
where the former intended use was by health care professionals only; or
    (2) The modified device operates using a different fundamental 
scientific technology than that in use in the device before May 28, 
1976; e.g., a surgical instrument cuts tissue with a laser beam rather 
than with a sharpened metal blade, or an in vitro diagnostic device 
detects or identifies infectious agents by using a deoxyribonucleic acid 
(DNA) probe or nucleic acid hybridization technology rather than culture 
or immunoassay technology.
    (b) The exemption from the requirement of premarket notification for 
a generic type of class II device applies only to those class II devices 
that have existing or reasonably foreseeable characteristics of 
commercially distributed devices within that generic type, or, in the 
case of in vitro diagnostic devices, for which a misdiagnosis, as a 
result of using the device, would not be associated with high morbidity 
or mortality. A class II device for which FDA has granted an exemption 
from the requirement of premarket notification must still submit a 
premarket notification when:
    (1) The device is intended for a use different from the intended use 
of a legally marketed device in that generic type of device; e.g., the 
device is intended for a different medical purpose, or the device is 
intended for lay use where the former intended use was by health care 
professionals only; or
    (2) The modified device operates using a different fundamental 
scientific technology than a legally marketed device in that generic 
type of device; e.g., a surgical instrument cuts tissue with a laser 
beam rather than with a sharpened metal blade, or an in vitro diagnostic 
device detects or identifies infectious agents by using deoxyribonucleic 
acid (DNA) probe or nucleic acid hybridization technology rather than 
culture or immunoassay technology; or
    (3) The device is an in vitro device that is intended:
    (i) For use in the diagnosis, monitoring, or screening of neoplastic 
diseases with the exception of immunohistochemical devices;
    (ii) For use in screening or diagnosis of familial and acquired 
genetic disorders, including inborn errors of metabolism;
    (iii) For measuring an analyte that serves as a surrogate marker for 
screening, diagnosis, or monitoring life-threatening diseases such as 
acquired immune deficiency syndrome

[[Page 480]]

(AIDS), chronic or active hepatitis, tuberculosis, or myocardial 
infarction or to monitor therapy;
    (iv) For assessing the risk of cardiovascular diseases;
    (v) For use in diabetes management;
    (vi) For identifying or inferring the identity of a microorganism 
directly from clinical material;
    (vii) For detection of antibodies to microorganisms other than 
immunoglobulin G (IgG) and IgG assays when the results are not 
qualitative, or are used to determine immunity, or the assay is intended 
for use in matrices other than serum or plasma;
    (viii) For noninvasive testing; and
    (ix) For near patient testing (point of care).

[54 FR 13831, Apr. 5, 1989, as amended at 63 FR 59231, Nov. 3, 1998]



                      Subpart B--Diagnostic Devices



Sec. 892.1000  Magnetic resonance diagnostic device.

    (a) Identification. A magnetic resonance diagnostic device is 
intended for general diagnostic use to present images which reflect the 
spatial distribution and/or magnetic resonance spectra which reflect 
frequency and distribution of nuclei exhibiting nuclear magnetic 
resonance. Other physical parameters derived from the images and/or 
spectra may also be produced. The device includes hydrogen-1 (proton) 
imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 
spectroscopy, and chemical shift imaging (preserving simultaneous 
frequency and spatial information).
    (b) Classification. Class II.

[53 FR 5078, Feb. 1, 1989]



Sec. 892.1100  Scintillation (gamma) camera.

    (a) Identification. A scintillation (gamma) camera is a device 
intended to image the distribution of radionuclides in the body by means 
of a photon radiation detector. This generic type of device may include 
signal analysis and display equipment, patient and equipment supports, 
radionuclide anatomical markers, component parts, and accessories.
    (b) Classification. Class I.

[55 FR 48443, Nov. 20, 1990]



Sec. 892.1110  Positron camera.

    (a) Identification. A positron camera is a device intended to image 
the distribution of positron-emitting radionuclides in the body. This 
generic type of device may include signal analysis and display 
equipment, patient and equipment supports, radionuclide anatomical 
markers, component parts, and accessories.
    (b) Classification. Class I.

[55 FR 48444, Nov. 20, 1990]



Sec. 892.1130  Nuclear whole body counter.

    (a) Identification. A nuclear whole body counter is a device 
intended to measure the amount of radionuclides in the entire body. This 
generic type of device may include signal analysis and display 
equipment, patient and equipment supports, component parts, and 
accessories.
    (b) Classification. Class I. The device is exempt from the premarket 
notification procedures in subpart E of part 807 of this chapter.

[53 FR 1567, Jan. 20, 1988, as amended at 59 FR 63015, Dec. 7, 1994]

[55 FR 48444, Nov. 20, 1990]



Sec. 892.1170  Bone densitometer.

    (a) Identification. A bone densitometer is a device intended for 
medical purposes to measure bone density and mineral content by x-ray or 
gamma ray transmission measurements through the bone and adjacent 
tissues. This generic type of device may include signal analysis and 
display equipment, patient and equipment supports, component parts, and 
accessories.
    (b) Classification. Class II.



Sec. 892.1200  Emission computed tomography system.

    (a) Identification. An emission computed tomography system is a 
device intended to detect the location and distribution of gamma ray- 
and positron-emitting radionuclides in the body and produce cross-
sectional images through computer reconstruction of the data.

[[Page 481]]

This generic type of device may include signal analysis and display 
equipment, patient and equipment supports, radionuclide anatomical 
markers, component parts, and accessories.
    (b) Classification. Class II.



Sec. 892.1220  Fluorescent scanner.

    (a) Identification. A fluorescent scanner is a device intended to 
measure the induced fluorescent radiation in the body by exposing the 
body to certain x-rays or low-energy gamma rays. This generic type of 
device may include signal analysis and display equipment, patient and 
equipment supports, component parts and accessories.
    (b) Classification. Class II.



Sec. 892.1300  Nuclear rectilinear scanner.

    (a) Identification. A nuclear rectilinear scanner is a device 
intended to image the distribution of radionuclides in the body by means 
of a detector (or detectors) whose position moves in two directions with 
respect to the patient. This generic type of device may include signal 
analysis and display equipment, patient and equipment supports, 
radionuclide anatomical markers, component parts, and accessories.
    (b) Classification. Class I.

[55 FR 48444, Nov. 20, 1990]



Sec. 892.1310  Nuclear tomography system.

    (a) Identification. A nuclear tomography system is a device intended 
to detect nuclear radiation in the body and produce images of a specific 
cross-sectional plane of the body by blurring or eliminating detail from 
other planes. This generic type of devices may include signal analysis 
and display equipment, patient and equipment supports, radionuclide 
anatomical markers, component parts, and accessories.
    (b) Classification. Class II.



Sec. 892.1320  Nuclear uptake probe.

    (a) Identification. A nuclear uptake probe is a device intended to 
measure the amount of radionuclide taken up by a particular organ or 
body region. This generic type of device may include a single or 
multiple detector probe, signal analysis and display equipment, patient 
and equipment supports, component parts, and accessories.
    (b) Classification. Class I.

[55 FR 48444, Nov. 20, 1990]



Sec. 892.1330  Nuclear whole body scanner.

    (a) Identification. A nuclear whole body scanner is a device 
intended to measure and image the distribution of radionuclides in the 
body by means of a wide-aperture detector whose position moves in one 
direction with respect to the patient. This generic type of device may 
include signal analysis and display equipment, patient and equipment 
supports, radionuclide anatomical markers, component parts, and 
accessories.
    (b) Classification. Class I.

[55 FR 48444, Nov. 20, 1990]



Sec. 892.1350  Nuclear scanning bed.

    (a) Identification. A nuclear scanning bed is an adjustable bed 
intended to support a patient during a nuclear medicine procedure.
    (b) Classification. Class I. The device is exempt from the premarket 
notification procedures in subpart E of part 807 of this chapter only 
when the device is labeled with weight limit, is used with planar 
scanning only, and is not for diagnostic X-ray use.

[55 FR 48444, Nov. 20, 1990, as amended at 59 FR 63015, Dec. 7, 1994]



Sec. 892.1360  Radionuclide dose calibrator.

    (a) Identification. A radionuclide dose calibrator is a radiation 
detection device intended to assay radionuclides before their 
administration to patients.
    (b) Classification. Class II.



Sec. 892.1370  Nuclear anthropomorphic phantom.

    (a) Identification. A nuclear anthropomorphic phantom is a human 
tissue facsimile that contains a radioactive source or a cavity in which 
a radioactive sample can be inserted. It is intended to calibrate 
nuclear uptake probes or other medical instruments.

[[Page 482]]

    (b) Classification. Class I. The device is exempt from the premarket 
notification procedures in subpart E of part 807 of this chapter.

[53 FR 1567, Jan. 20, 1988, as amended at 54 FR 13832, Apr. 5, 1989]



Sec. 892.1380  Nuclear flood source phantom.

    (a) Identification. A nuclear flood source phantom is a device that 
consists of a radiolucent container filled with a uniformly distributed 
solution of a desired radionuclide. It is intended to calibrate a 
medical gamma camera-collimator system for uniformity of response.
    (b) Classification. Class I. The device is exempt from the premarket 
notification procedures in subpart E of part 807 of this chapter.

[53 FR 1567, Jan. 20, 1988, as amended at 54 FR 13832, Apr. 5, 1989]



Sec. 892.1390  Radionuclide rebreathing system.

    (a) Identification. A radionuclide rebreathing system is a device 
intended to be used to contain a gaseous or volatile radionuclide or a 
radionuclide-labeled aerosol and permit it to be respired by the patient 
during nuclear medicine ventilatory tests (testing process of exchange 
between the lungs and the atmosphere). This generic type of device may 
include signal analysis and display equipment, patient and equipment 
supports, component parts, and accessories.
    (b) Classification. Class II.



Sec. 892.1400  Nuclear sealed calibration source.

    (a) Identification. A nuclear sealed calibration source is a device 
that consists of an encapsulated reference radionuclide intended for 
calibration of medical nuclear radiation detectors.
    (b) Classification. Class I. The device is exempt from the premarket 
notification procedures in subpart E of part 807 of this chapter.

[53 FR 1567, Jan. 20, 1988, as amended at 54 FR 13832, Apr. 5, 1989]



Sec. 892.1410  Nuclear electrocardiograph synchronizer.

    (a) Identification. A nuclear electrocardiograph synchronizer is a 
device intended for use in nuclear radiology to relate the time of image 
formation to the cardiac cycle during the production of dynamic cardiac 
images.
    (b) Classification. Class I.

[55 FR 48444, Nov. 20, 1990]



Sec. 892.1420  Radionuclide test pattern phantom.

    (a) Identification. A radionuclide test pattern phantom is a device 
that consists of an arrangement of radiopaque or radioactive material 
sealed in a solid pattern intended to serve as a test for a performance 
characteristic of a nuclear medicine imaging device.
    (b) Classification. Class I. The device is exempt from the premarket 
notification procedures in subpart E of part 807 of this chapter.

[53 FR 1567, Jan. 20, 1988, as amended at 54 FR 13832, Apr. 5, 1989]



Sec. 892.1540  Nonfetal ultrasonic monitor.

    (a) Identification. A nonfetal ultrasonic monitor is a device that 
projects a continuous high-frequency sound wave into body tissue other 
than a fetus to determine frequency changes (doppler shift) in the 
reflected wave and is intended for use in the investigation of nonfetal 
blood flow and other nonfetal body tissues in motion. This generic type 
of device may include signal analysis and display equipment, patient and 
equipment supports, component parts, and accessories.
    (b) Classification. Class II.



Sec. 892.1550  Ultrasonic pulsed doppler imaging system.

    (a) Identification. An ultrasonic pulsed doppler imaging system is a 
device that combines the features of continuous wave doppler-effect 
technology with pulsed-echo effect technology and is intended to 
determine stationary body tissue characteristics, such as depth or 
location of tissue interfaces or dynamic tissue characteristics such as 
velocity of blood or tissue motion. This

[[Page 483]]

generic type of device may include signal analysis and display 
equipment, patient and equipment supports, component parts, and 
accessories.
    (b) Classification. Class II.



Sec. 892.1560  Ultrasonic pulsed echo imaging system.

    (a) Identification. An ultrasonic pulsed echo imaging system is a 
device intended to project a pulsed sound beam into body tissue to 
determine the depth or location of the tissue interfaces and to measure 
the duration of an acoustic pulse from the transmitter to the tissue 
interface and back to the receiver. This generic type of device may 
include signal analysis and display equipment, patient and equipment 
supports, component parts, and accessories.
    (b) Classification. Class II.



Sec. 892.1570  Diagnostic ultrasonic transducer.

    (a) Identification. A diagnostic ultrasonic transducer is a device 
made of a piezoelectric material that converts electrical signals into 
acoustic signals and acoustic signals into electrical signals and 
intended for use in diagnostic ultrasonic medical devices. Accessories 
of this generic type of device may include transmission media for 
acoustically coupling the transducer to the body surface, such as 
acoustic gel, paste, or a flexible fluid container.
    (b) Classification. Class II.



Sec. 892.1600  Angiographic x-ray system.

    (a) Identification. An angiographic x-ray system is a device 
intended for radiologic visualization of the heart, blood vessels, or 
lymphatic system during or after injection of a contrast medium. This 
generic type of device may include signal analysis and display 
equipment, patient and equipment supports, component parts, and 
accessories.
    (b) Classification. Class II.



Sec. 892.1610  Diagnostic x-ray beam-limiting device.

    (a) Identification. A diagnostic x-ray beam-limiting device is a 
device such as a collimator, a cone, or an aperture intended to restrict 
the dimensions of a diagnostic x-ray field by limiting the size of the 
primary x-ray beam.
    (b) Classification. Class II.



Sec. 892.1620  Cine or spot fluorographic x-ray camera.

    (a) Identification. A cine or spot fluorographic x-ray camera is a 
device intended to photograph diagnostic images produced by x-rays with 
an image intensifier.
    (b) Classification. Class II.



Sec. 892.1630  Electrostatic x-ray imaging system.

    (a) Identification. An electrostatic x-ray imaging system is a 
device intended for medical purposes that uses an electrostatic field 
across a semiconductive plate, a gas-filled chamber, or other similar 
device to convert a pattern of x-radiation into an electrostatic image 
and, subsequently, into a visible image. This generic type of device may 
include signal analysis and display equipment, patient and equipment 
supports, component parts, and accessories.
    (b) Classification. Class II.



Sec. 892.1640  Radiographic film marking system.

    (a) Identification. A radiographic film marking system is a device 
intended for medical purposes to add identification and other 
information onto radiographic film by means of exposure to visible 
light.
    (b) Classification. Class I. The device is exempt from the premarket 
notification procedures in subpart E of part 807 of this chapter.

[55 FR 48444, Nov. 20, 1990, as amended at 59 FR 63015, Dec. 7, 1994]



Sec. 892.1650  Image-intensified fluoroscopic x-ray system.

    (a) Identification. An image-intensified fluoroscopic x-ray system 
is a device intended to visualize anatomical structures by converting a 
pattern of x-radiation into a visible image through electronic 
amplification. This generic type of device may include signal analysis 
and display equipment, patient and equipment supports, component parts, 
and accessories.
    (b) Classification. Class II.

[[Page 484]]



Sec. 892.1660  Non-image-intensified fluoroscopic x-ray system.

    (a) Identification. A non-image-intensified fluoroscopic x-ray 
system is a device intended to be used to visualize anatomical 
structures by using a fluorescent screen to convert a pattern of x-
radiation into a visible image. This generic type of device may include 
signal analysis and display equipment, patient and equipment supports, 
component parts, and accessories.
    (b) Classification. Class II.



Sec. 892.1670  Spot-film device.

    (a) Identification. A spot-film device is an electromechanical 
component of a fluoroscopic x-ray system that is intended to be used for 
medical purposes to position a radiographic film cassette to obtain 
radiographs during fluoroscopy.
    (b) Classification. Class II.



Sec. 892.1680  Stationary x-ray system.

    (a) Identification. A stationary x-ray system is a permanently 
installed diagnostic system intended to generate and control x-rays for 
examination of various anatomical regions. This generic type of device 
may include signal analysis and display equipment, patient and equipment 
supports, component parts, and accessories.
    (b) Classification. Class II.



Sec. 892.1700  Diagnostic x-ray high voltage generator.

    (a) Identification. A diagnostic x-ray high voltage generator is a 
device that is intended to supply and control the electrical energy 
applied to a diagnostic x-ray tube for medical purposes. This generic 
type of device may include a converter that changes alternating current 
to direct current, filament transformers for the x-ray tube, high 
voltage switches, electrical protective devices, or other appropriate 
elements.
    (b) Classification. Class I. The device is exempt from the premarket 
notification procedures in subpart E of part 807 of this chapter.

[53 FR 1567, Jan. 20, 1988, as amended at 61 FR 1125, Jan. 16, 1996]



Sec. 892.1710  Mammographic x-ray system.

    (a) Identification. A mammographic x-ray system is a device intended 
to be used to produce radiographs of the breast. This generic type of 
device may include signal analysis and display equipment, patient and 
equipment supports, component parts, and accessories.
    (b) Classification. Class II.



Sec. 892.1720  Mobile x-ray system.

    (a) Identification. A mobile x-ray system is a transportable device 
system intended to be used to generate and control x-ray for diagnostic 
procedures. This generic type of device may include signal analysis and 
display equipment, patient and equipment supports, component parts, and 
accessories.
    (b) Classification. Class II.



Sec. 892.1730  Photofluorographic x-ray system.

    (a) Identification. A photofluorographic x-ray system is a device 
that includes a fluoroscopic x-ray unit and a camera intended to be used 
to produce, then photograph, a fluoroscopic image of the body. This 
generic type of device may include signal analysis and display 
equipment, patient and equipment supports, component parts, and 
accessories.
    (b) Classification. Class II.



Sec. 892.1740  Tomographic x-ray system.

    (a) Identification. A tomographic x-ray system is an x-ray device 
intended to be used to produce radiologic images of a specific cross-
sectional plane of the body by blurring or eliminating detail from other 
planes. This generic type of device may include signal analysis and 
display equipment, patient and equipment supports, component parts, and 
accessories.
    (b) Classification. Class II.



Sec. 892.1750  Computed tomography x-ray system.

    (a) Identification. A computed tomography x-ray system is a 
diagnostic x-ray system intended to produce cross-sectional images of 
the body by computer reconstruction of x-ray transmission data from the 
same axial plane taken at different angles. This generic

[[Page 485]]

type of device may include signal analysis and display equipment, 
patient and equipment supports, component parts, and accessories.
    (b) Classification. Class II.



Sec. 892.1760  Diagnostic x-ray tube housing assembly.

    (a) Identification. A diagnostic x-ray tube housing assembly is an 
x-ray generating tube encased in a radiation-shielded housing that is 
intended for diagnostic purposes. This generic type of device may 
include high voltage and filament transformers or other appropriate 
components.
    (b) Classification. Class I. The device is exempt from the premarket 
notification procedures in subpart E of part 807 of this chapter.

[53 FR 1567, Jan. 20, 1988, as amended at 61 FR 1125, Jan. 16, 1996]



Sec. 892.1770  Diagnostic x-ray tube mount.

    (a) Identification. A diagnostic x-ray tube mount is a device 
intended to support and to position the diagnostic x-ray tube housing 
assembly for a medical radiographic procedure.
    (b) Classification. Class I. The device is exempt from the premarket 
notification procedures in subpart E of part 807 of this chapter.

[53 FR 1567, Jan. 20, 1988, as amended at 61 FR 1125, Jan. 16, 1996]



Sec. 892.1820  Pneumoencephalographic chair.

    (a) Identification. A pneumoencephalographic chair is a chair 
intended to support and position a patient during pneumoencephalography 
(x-ray imaging of the brain).
    (b) Classification. Class II.



Sec. 892.1830  Radiologic patient cradle.

    (a) Identification. A radiologic patient cradle is a support device 
intended to be used for rotational positioning about the longitudinal 
axis of a patient during radiologic procedures.
    (b) Classification. Class I. The device is exempt from the premarket 
notification procedures in subpart E of part 807 of this chapter.

[53 FR 1567, Jan. 20, 1988, as amended at 61 FR 1125, Jan. 16, 1996]



Sec. 892.1840  Radiographic film.

    (a) Identification. Radiographic film is a device that consists of a 
thin sheet of radiotransparent material coated on one or both sides with 
a photographic emulsion intended to record images during diagnostic 
radiologic procedures.
    (b) Classification. Class I. The device is exempt from the premarket 
notification procedures in subpart E of part 807.



Sec. 892.1850  Radiographic film cassette.

    (a) Identification. A radiographic film cassette is a device 
intended for use during diagnostic x-ray procedures to hold a 
radiographic film in close contact with an x-ray intensifying screen and 
to provide a light-proof enclosure for direct exposure of radiographic 
film.
    (b) Classification. Class II.



Sec. 892.1860  Radiographic film/cassette changer.

    (a) Identification. A radiographic film/cassette changer is a device 
intended to be used during a radiologic procedure to move a radiographic 
film or cassette between x-ray exposures and to position it during the 
exposure.
    (b) Classification. Class II.



Sec. 892.1870  Radiographic film/cassette changer programmer.

    (a) Identification. A radiographic film/cassette changer programmer 
is a device intended to be used to control the operations of a film or 
cassette changer during serial medical radiography.
    (b) Classification. Class II.



Sec. 892.1880  Wall-mounted radiographic cassette holder.

    (a) Identification. A wall-mounted radiographic cassette holder is a 
device that is a support intended to hold and position radiographic 
cassettes for a radiographic exposure for medical use.
    (b) Classification. Class I. The device is exempt from the premarket 
notification procedures in subpart E of part 807 of this chapter.

[53 FR 1567, Jan. 20, 1988, as amended at 61 FR 1125, Jan. 16, 1996]

[[Page 486]]



Sec. 892.1890  Radiographic film illuminator.

    (a) Identification. A radiographic film illuminator is a device 
containing a visible light source covered with a translucent front that 
is intended to be used to view medical radiographs.
    (b) Classification. Class I.

[55 FR 48444, Nov. 20, 1990]



Sec. 892.1900  Automatic radiographic film processor.

    (a) Identification. An automatic radiographic film processor is a 
device intended to be used to develop, fix, wash, and dry automatically 
and continuously film exposed for medical purposes.
    (b) Classification. Class II.

[55 FR 48444, Nov. 20, 1990]



Sec. 892.1910  Radiographic grid.

    (a) Identification. A radiographic grid is a device that consists of 
alternating radiolucent and radiopaque strips intended to be placed 
between the patient and the image receptor to reduce the amount of 
scattered radiation reaching the image receptor.
    (b) Classification. Class I.



Sec. 892.1920  Radiographic head holder.

    (a) Identification. A radiographic head holder is a device intended 
to position the patient's head during a radiographic procedure.
    (b) Classification. Class I. The device is exempt from the premarket 
notification procedures in subpart E of part 807. The device is exempt 
from the current good manufacturing practice regulations in part 820, 
with the exception of Sec. 820.180, with respect to general requirements 
concerning records, and Sec. 820.198, with respect to complaint files.



Sec. 892.1940  Radiologic quality assurance instrument.

    (a) Identification. A radiologic quality assurance instrument is a 
device intended for medical purposes to measure a physical 
characteristic associated with another radiologic device.
    (b) Classification. Class I. The device is exempt from the premarket 
notification procedures in subpart E of part 807. The device is exempt 
from the current good manufacturing practice regulations in part 820, 
with the exception of Sec. 820.180, with respect to general requirements 
concerning records, and Sec. 820.198, with respect to complaint files.



Sec. 892.1950  Radiographic anthropomorphic phantom.

    (a) Identification. A radiographic anthropomorphic phantom is a 
device intended for medical purposes to simulate a human body for 
positioning radiographic equipment.
    (b) Classification. Class I. The device is exempt from the premarket 
notification procedures in subpart E of part 807. The device is exempt 
from the current good manufacturing practice regulations in part 820, 
with the exception of Sec. 820.180, with respect to general requirements 
concerning records, and Sec. 820.198, with respect to complaint files.



Sec. 892.1960  Radiographic intensifying screen.

    (a) Identification. A radiographic intensifying screen is a device 
that is a thin radiolucent sheet coated with a luminescent material that 
transforms incident x-ray photons into visible light and intended for 
medical purposes to expose radiographic film.
    (b) Classification. Class I.



Sec. 892.1970  Radiographic ECG/respirator synchronizer.

    (a) Identification. A radiographic ECG/respirator synchronizer is a 
device intended to be used to  coordinate  an x-ray film exposure with 
the signal from an electrocardiograph (ECG) or respirator at a 
predetermined phase of the cardiac or respiratory cycle.
    (b) Classification. Class I.

[55 FR 48444, Nov. 20, 1990]



Sec. 892.1980  Radiologic table.

    (a) Identification. A radiologic table is a device intended for 
medical purposes to support a patient during radiologic procedures. The 
table may be fixed or tilting and may be electrically powered.
    (b) Classification. Class II (special controls). The device is 
exempt from the premarket notification procedures

[[Page 487]]

in subpart E of part 807 of this chapter subject to Sec. 892.9.

[53 FR 1567, Jan. 20, 1988, as amended at 63 FR 59231, Nov. 3, 1998]



Sec. 892.1990  Transilluminator for breast evaluation.

    (a) Identification. A transilluminator, also known as a 
diaphanoscope or lightscanner, is an electrically powered device that 
uses low intensity emissions of visible light and near-infrared 
radiation (approximately 700-1050 nanometers (nm)), transmitted through 
the breast, to visualize translucent tissue for the diagnosis of cancer, 
other conditions, diseases, or abnormalities.
    (b) Classification. Class III (premarket approval).
    (c) Date premarket approval (PMA) or notice of completion of a 
product development protocol (PDP) is required. The effective date of 
the requirement for premarket approval has not been established. See 
Sec. 892.3.

[60 FR 36639, July 18, 1995]



Sec. 892.2010  Medical image storage device.

    (a) Identification. A medical image storage device is a device that 
provides electronic storage and retrieval functions for medical images. 
Examples include devices employing magnetic and optical discs, magnetic 
tape, and digital memory.
    (b) Classification. Class I. The device is exempt from the premarket 
notification procedures in subpart E of part 807 of this chapter only 
when the device stores images without performing irreversible data 
compression.

[63 FR 23387, Apr. 29, 1998; 63 FR 44998, Aug. 24, 1998]



Sec. 892.2020  Medical image communications device.

    (a) Identification. A medical image communications device provides 
electronic transfer of medical image data between medical devices. It 
may include a physical communications medium, modems, interfaces, and a 
communications protocol.
    (b) Classification. Class I. The device is exempt from the premarket 
notification procedures in subpart E of part 807 of this chapter only 
when the device transfers images without performing irreversible data 
compression.

[63 FR 23387, Apr. 29, 1998; 63 FR 44998, Aug. 24, 1998]



Sec. 892.2030  Medical image digitizer.

    (a) Identification. A medical image digitizer is a device intended 
to convert an analog medical image into a digital format. Examples 
include Iystems employing video frame grabbers, and scanners which use 
lasers or charge-coupled devices.
    (b) Classification. Class II (special controls; voluntary 
standards--Digital Imaging and Communications in Medicine (DICOM) Std., 
Joint Photographic Experts Group (JPEG) Std.).

[63 FR 23387, Apr. 29, 1998]



Sec. 892.2040  Medical image hardcopy device.

    (a) Identification. A medical image hardcopy device is a device that 
produces a visible printed record of a medical image and associated 
identification information. Examples include multiformat cameras and 
laser printers.
    (b) Classification. Class II (special controls; voluntary 
standards--Digital Imaging and Communications in Medicine (DICOM) Std., 
Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture 
and Television Engineers (SMPTE) Test Pattern).

[63 FR 23387, Apr. 29, 1998]



Sec. 892.2050  Picture archiving and communications system.

    (a) Identification. A picture archiving and communications system is 
a device that provides one or more capabilities relating to the 
acceptance, transfer, display, storage, and digital processing of 
medical images. Its hardware components may include workstations, 
digitizers, communications devices, computers, video monitors, magnetic, 
optical disk, or other digital data storage devices, and hardcopy 
devices. The software components may provide functions for performing 
operations related to image manipulation, enhancement, compression or 
quantification.
    (b) Classification. Class II (special controls; voluntary 
standards--Digital

[[Page 488]]

Imaging and Communications in Medicine (DICOM) Std., Joint Photographic 
Experts Group (JPEG) Std., Society of Motion Picture and Television 
Engineers (SMPTE) Test Pattern).

[63 FR 23387, Apr. 29, 1998]



                        Subparts C-E  [Reserved]



                     Subpart F--Therapeutic Devices



Sec. 892.5050  Medical charged-particle radiation therapy system.

    (a) Identification. A medical charged-particle radiation therapy 
system is a device that produces by acceleration high energy charged 
particles (e.g., electrons and protons) intended for use in radiation 
therapy. This generic type of device may include signal analysis and 
display equipment, patient and equipment supports, treatment planning 
computer programs, component parts, and accessories.
    (b) Classification. Class II. When intended for use as a quality 
control system, the film dosimetry system (film scanning system) 
included as an accessory to the device described in paragraph (a) of 
this section, is exempt from the premarket notification procedures in 
subpart E of part 807 of this chapter subject to the limitations in 
Sec. 892.9.

[53 FR 1567, Jan. 20, 1988, as amended at 64 FR 1125, Jan. 8, 1999]



Sec. 892.5300  Medical neutron radiation therapy system.

    (a) Identification. A medical neutron radiation therapy system is a 
device intended to generate high-energy neutrons for radiation therapy. 
This generic type of device may include signal analysis and display 
equipment, patient and equipment support, treatment planning computer 
programs, component parts, and accessories.
    (b) Classification. Class II.



Sec. 892.5650  Manual radionuclide applicator system.

    (a) Identification. A manual radionuclide applicator system is a 
manually operated device intended to apply a radionuclide source into 
the body or to the surface of the body for radiation therapy. This 
generic type of device may include patient and equipment supports, 
component parts, treatment planning computer programs, and accessories.
    (b) Classification. Class I.



Sec. 892.5700  Remote controlled radionuclide applicator system.

    (a) Identification. A remote controlled radionuclide applicator 
system is an electromechanical or pneumatic device intended to enable an 
operator to apply, by remote control, a radionuclide source into the 
body or to the surface of the body for radiation therapy. This generic 
type of device may include patient and equipment supports, component 
parts, treatment planning computer programs, and accessories.
    (b) Classification. Class II.



Sec. 892.5710  Radiation therapy beam-shaping block.

    (a) Identification. A radiation therapy beam-shaping block is a 
device made of a highly attenuating material (such as lead) intended for 
medical purposes to modify the shape of a beam from a radiation therapy 
source.
    (b) Classification. Class II.



Sec. 892.5730  Radionuclide brachytherapy source.

    (a) Identification. A radionuclide brachytherapy source is a device 
that consists of a radionuclide which may be enclosed in a sealed 
container made of gold, titanium, stainless steel, or platinum and 
intended for medical purposes to be placed onto a body surface or into a 
body cavity or tissue as a source of nuclear radiation for therapy.
    (b) Classification. Class II.



Sec. 892.5740  Radionuclide teletherapy source.

    (a) Identification. A radionuclide teletherapy source is a device 
consisting of a radionuclide enclosed in a sealed container. The device 
is intended for radiation therapy, with the radiation source located at 
a distance from the patient's body.

[[Page 489]]

    (b) Classification. Class I. The device is exempt from the premarket 
notification procedures in subpart E of part 807 of this chapter.

[53 FR 1567, Jan. 20, 1988, as amended at 59 FR 63015, Dec. 7, 1994]



Sec. 892.5750  Radionuclide radiation therapy system.

    (a) Identification. A radionuclide radiation therapy system is a 
device intended to permit an operator to administer gamma radiation 
therapy, with the radiation source located at a distance from the 
patient's body. This generic type of device may include signal analysis 
and display equipment, patient and equipment supports, treatment 
planning computer programs, component parts (including beam-limiting 
devices), and accessories.
    (b) Classification. Class II.



Sec. 892.5770  Powered radiation therapy patient support assembly.

    (a) Identification. A powered radiation therapy patient support 
assembly is an electrically powered adjustable couch intended to support 
a patient during radiation therapy.
    (b) Classification. Class II.



Sec. 892.5780  Light beam patient position indicator.

    (a) Identification. A light beam patient position indicator is a 
device that projects a beam of light (incoherent light or laser) to 
determine the alignment of the patient with a radiation beam. The beam 
of light is intended to be used during radiologic procedures to ensure 
proper positioning of the patient and to monitor alignment of the 
radiation beam with the patient's anatomy.
    (b) Classification. Class I. The device is exempt from the premarket 
notification procedures in subpart E of part 807 of this chapter.

[53 FR 1567, Jan. 20, 1988, as amended at 61 FR 1125, Jan. 16, 1996]



Sec. 892.5840  Radiation therapy simulation system.

    (a) Identification. A radiation therapy simulation system is a 
fluoroscopic or radiographic x-ray system intended for use in localizing 
the volume to be exposed during radiation therapy and confirming the 
position and size of the therapeutic irradiation field produced. This 
generic type of device may include signal analysis and display 
equipment, patient and equipment supports, treatment planning computer 
programs, component parts, and accessories.
    (b) Classification. Class II.



Sec. 892.5900  X-ray radiation therapy system.

    (a) Identification. An x-ray radiation therapy system is a device 
intended to produce and control x-rays used for radiation therapy. This 
generic type of device may include signal analysis and display 
equipment, patient and equipment supports, treatment planning computer 
programs, component parts, and accessories.
    (b) Classification. Class II.



Sec. 892.5930  Therapeutic x-ray tube housing assembly.

    (a) Identification. A therapeutic x-ray tube housing assembly is an 
x-ray generating tube encased in a radiation-shielded housing intended 
for use in radiation therapy. This generic type of device may include 
high-voltage and filament transformers or other appropriate components 
when contained in radiation-shielded housing.
    (b) Classification. Class II.



                    Subpart G--Miscellaneous Devices



Sec. 892.6500  Personnel protective shield.

    (a) Identification. A personnel protective shield is a device 
intended for medical purposes to protect the patient, the operator, or 
other persons from unnecessary exposure to radiation during radiologic 
procedures by providing an attenuating barrier to radiation. This 
generic type of device may include articles of clothing, furniture, and 
movable or stationary structures.
    (b) Classification. Class I. If the device's labeling specifies the 
lead equivalence, it is exempt from the premarket notification 
procedures in subpart E of part 807 of this chapter.

[53 FR 1567, Jan. 20, 1988, as amended at 61 FR 1125, Jan. 16, 1996]

[[Page 490]]